Bender 107 TL 47 Insulation Monitoring Device
Insulation Monitoring Device
Insulation monitoring device with load current and temperature monitoring, designed for IT power systems in medical locations in compliance with international safety standards.BENDER 107TL47 Insulation Monitoring Device
The BENDER 107TL47 is an advanced insulation monitoring device designed for IT systems in medical locations. It provides continuous insulation monitoring with integrated load current and temperature monitoring, ensuring maximum safety and compliance with medical electrical standards.
Manufacturer: BENDER
Type: Insulation Monitoring Device with Load Current and Temperature Monitoring
Model: 107 TL 47
Art. No.: 922178
Application:
IT systems for power supply in medical locations, hospitals, medical practices, and outpatient surgical centers
IT systems with restricted power source with additional load current monitoring
| Range | ISOMETER® |
| Series | 107TD47 |
| Used For | IT Systems in Medical Locations According to DIN VDE 0100-710 and IEC 60364-7-710 |
| Rated Insulation Voltage | 250 V AC |
| Rated Impulse Voltage | 4 kV |
| Frequency | 40 / 460 Hz |
| Power Consumption (Max.) | 3 VA |
| Current Measuring Range | 5 / 50 A |
| Response Value | 50 / 500 kΩ |
| Interface / Protocol | RS-485 / BMS |
| Contact Switching Elements | 1 Changeover (CO) |
| Protection Class | IP30 |
| Ambient Temperature | -10 / 55 °C |
| Dimensions | 73 x 99 x 75 mm |